Micro-scale VOC Emission Chamber

Inquire product
  • Adjustable range of temperature: 60~200℃, High temperature cleaning 260℃
  • Adjustable range of humidity: 5%R.H.(65℃)
  • Air exchange rate: 10-500time/h±5%
  • Air velocity: 0.1m/s~0.3m/s(Range)
  • Background concentration of VOC: Single voc concentration is smaller than 2μg/m³,concentration is smaller than 20μg/m³
  • Rate of recovery: Formaldehyde, octanol, 4 phenyl Cyclohexene, dodecane and tetradecane have a recovery rate more than 80%
  • Reference Standards: ISO 12219-3

FUNCTION

Micro sampling chamber can provide standard testing environment(temperature,cleanliness,relative humility,air exchange rate,etc),and detect VOC/SVOC and water content of various materials and products.It is very suitable for concrete analysis of teterogeneous prototype and detection of low trace elements.

Reference Standards

ISO 12219-3:2012(E)air in road vehicles Section 3 Determination Method for screening release emission of volatile organic components of materials of vehicle interiors with micro test chamber method

TECHICAL PARAMETERS

Product ModelMV-4; MV-6; MV-8;
Environmental temperature5℃~35℃
Environmental hnmidity≤90%RH
Requirements for EnvironmentalNo pollutant source surrounding TVOC≤500μg/m³
Volume of climate chamberφ64xh34mm=116ml,Inner arc surface chamber
Quantity of Inner tanksseveral,with aspiration at the bottom ring and outlet and sampling at the ceiling
Adjustable range of temperature60~200℃,High temperature cleaning 260℃
Adjustable range of humidity(optional)5%R.H.(65℃)
PressurePositive pressure
Temperature fluctuation±0.3℃
Temperature deviation±1.0℃
Humidity fluctuation(optional)±2.5%R.H.
Humidity deviation(optional)±3%
Air exchange rate(optional)10-500time/h±5%
Air flow volume10-500mL/min(Quality flow controller)
PowerAC 220V/50Hz
Starting power3kw
Operating power1.5kw
Outline dimension of the climate chamberW580×D890×H1050mm(when there are 6 Inner tanks),which shall be confirmed according to the quantity of inner tanks.
Background concentration of formaldehydeSmaller than 15μg/m³,the lowest  6μg/ m³
Background concentration of VOCSingle voc concentration is smaller than 2μg/m³,concentration is smaller than 20μg/m³
Rate of recoveryFormaldehyde,octanol,4 phenyl Cyclohexene, dodecane and tetradecane have a recovery rate more than 80%
Air purification system dry and clean gases(Non combustible gases and Non toxic and harmful gases such as air,nitrogen and argon) pressure>1.0Mpa
Cooling modeAir cooling
ControllerColor touch screen man-machine interface Mitsubishi PLC controller