Incubated Shakers Chambers

Inquire product

Specially designed to be double-stacked on the floor or located on the lab benches for space saving with optional accessories

Structural Functional Features

  • Best space-efficient model with double stacking.
  • Up to 80°C, 500rpm enables setting of various experimental conditions.
  • Three operating modes: Temperature & Shaking, Temperature, or Shaking.
  • Includes internal power outlet.
  • Shaking structure with minimized vibration.

Use Convenience Features

  • Accurate rpm and temperature control through temperature auto-tuning / microprocessor PID and calibration.
  • Monitor via mobile app anytime, anywhere with LC Connected. (mobile monitoring system) (when purchased LC GreenBox)
  • Countdown timer (999 hours 59 minutes) of shaking operation.
  • Intuitive operation with color touch display.
  • High brightness / low-power LED illumination makes it easy to observe inside.
  • Spill drain port for internal cleanliness and easy maintenance.

Outstanding Safety

  • Electronic over temperature protection system. (advanced over temperature (A-OT) limiter)
    Independent precision temperature sensor and controller provides more secure and reliable over temperature protection.
  • Conventional over-temperature protection system. (B-OT, backup over temperature limiter)
    A backup device against electronic system failure, mechanically preventing over temperature even when there are electronic errors.
  • Soft start and stop shaking function.
  • Malfunction prevented by controller lock function.
  • Displays all information such as device operation and error occurrence.
  • Overcurrent protection device restricts operation when objects are caught inside the system and shaking is not possible.

Smart Self-diagnosis

Real-time self-diagnostics to check equipment for abnormal conditions.

  • Alarm is generated by monitoring difference between actual rpm/temperature and set rpm/temperature.
  • Detects abnormal conditions relating to the platform on which the sample is placed and stops the operation.
  • Automatic restart as it was operated after power recovery.
Technical Specifications
Model ISS-3075 ISS-3075R ISS-4075 ISS-4075R
Temperature Range (℃ / ℉)1) Amb. +5 to 80 

/ Amb. +9 to 176

Amb. -20 (Min. 4) to 80 

/Amb. -36 (Min. 7.2) to 176

Amb. +5 to 80 

/Amb. +9 to 176

Amb. -20 (Min. 4) to 80 

/Amb. -36 (Min. 7.2) to 176

Fluctuation at 37℃ in flask 

(±℃ / ℉)

0.1 / 0.18
Variation at 37℃ in flask 

(±℃ / ℉)

0.5 / 0.90
Refrigerator (Hp) - 1/6Hp - 1/6Hp
Shaking system Motion type Orbital
Amplitude size (mm / inch, dia.) 19.1 / 0.75
Speed range (rpm) 20 to 500 (stackable : 20 to 250)
Speed stability2)(rpm, %) ±1
Timer (count-down) 999 hr 59 min.
Max. Speed per permissible Load (rpm) 500 (max.10kg / 22.0 lbs) 

400 (max.15kg / 33.1 lbs)

250 (max.15kg / 33.1 lbs) - (stacked)

500 (max.10kg / 22.0 lbs) 

400 (max.21kg / 46.3 lbs)

250 (max.21kg / 46.3 lbs) - (stacked)

Dimension Volume (L / cu ft) 80 / 2.8 150 / 5.3
Platform (W×D, mm / inch) 350×350 / 13.8x13.8 450×450 / 17.7×17.7
Interior (W×D×H, mm / inch) 440×440×418 / 17.3×17.3×16.5 540×540×518 / 21.3×21.3×20.4
Exterior (W×D×H, mm / inch) 578×815×854 / 22.8x32.1x33.6 678×895×934 / 26.7x35.2x36.8
Shelves (Standard / Max) 2 / 8 2 / 11
Net weight (kg / Ibs) 153 / 337.3 160 / 352.7 183 / 403.4 190 / 418.9
Electrical requirements (230V, 50/60Hz, A ) 6.2 7.9 6.2 7.9
Cat. No. AAH23432K AAH23532K AAH23632K AAH23732K
Electrical requirements (120V, 60Hz, A) 9.5 13 9.5 13
Cat. No. AAH23415U AAH23515U AAH23615U AAH23715U

1) The temperature control range of the IST-3075 and IST-4075 models are from 5°C above the ambient temperature of the equipment. (15°C to 40°C) to 80°C.
The temperature control range of the IST-3075R and IST-4075R models are from 15°C below the ambient temperature (18°C to 35°C) of the device to 80°C. (However, the lowest control temperature is 15°C)

2) Actual rpm allowable error for set rpm is ±1rpm if setting is less than 100rpm, ±1% if setting is higher than 100rpm.